Anritsu 0701012 User Manual
Page 40

Remote Operation Command Details
Remote Operation
4-4
MA24106A UG
HAT
Description:
Sets the high aperture time mode.
Syntax:
HAT + LF
Return Value:
OK or ERR
Remarks:
This command will put the sensor in high aperture time mode. In this mode, the A to D
converter integration time is about 160 milliseconds.
LAT
Description:
Sets the low aperture time mode.
Syntax:
HAT + LF
Return Value:
OK or ERR
Remarks:
This command will put the sensor in low aperture time mode. In this mode, the A to D
converter integration time is about 10 milliseconds. This mode is the default mode for the
sensor when powered up.