Sensor Operational Tests
Linearity Test
MA24106A UG
5-9
Table 5-5.
Measurement Results (2 GHz)
Row
#
Synthesizer
Power Level
Setting
(dBm)
Attenuation
in Reference
Arm
(dB)
A
Reference
Power
Measurement
(dBm)
B
= (A6–A5)
Correction
(dB)
C
= (A+B)
Corrected
Reference Power
Measurement
(dB)
Attenuation
in Test Arm
(dB)
D
MA24106A
Measurement
(dBm)
C–D
Difference
Calculation
(dB)
1
+20
10
0
2
+15
10
0
3
+10
10
0
4
+5
10
0
5
0
10
0
6
adjust per
step 7e
0
0
10
7
+5
0
0
10
8
0
0
0
10
9
–5
0
0
10
10
–10
0
0
10
11
Max:
12
Min:
13
Delta (E11 – E12):
Table 5-6.
Measurement Results (4 GHz)
Row
#
Synthesizer
Power Level
Setting
(dBm)
Attenuation
in Reference
Arm
(dB)
A
Reference
Power
Measurement
(dBm)
B
= (A6–A5)
Correction
(dB)
C
= (A+B)
Corrected
Reference Power
Measurement
(dB)
Attenuation
in Test Arm
(dB)
D
MA24106A
Measurement
(dBm)
E
= (C–D)
Difference
Calculation
(dB)
1
+20
10
0
2
+15
10
0
3
+10
10
0
4
+5
10
0
5
0
10
0
6
adjust per
step 7e
0
0
10
7
+5
0
0
10
8
0
0
0
10
9
–5
0
0
10
10
–10
0
0
10
11
Max:
12
Min:
13
Delta (E11 – E12):