1. Set switches 1, 2, 3, 4, 5 and 8 to OFF.
2. Set switches 6, 7, 9 and 10 to ON. Switch 9 and 10 set to ON sets a

1200 baud rate.

3. Refer to Setting of Preset Jumper in the Citizen Printer Manual.

Figure 10 Citizen printer switch configuration

Table 4 Dip switch settings for the Citizen printer

USA

Germany France

UK

Foreign character selection

DSW1-4

Off

Off

On

On

DSW1-5

Off

On

Off

On

Odd

Even

No Check

Parity check

DSW1-6

Off

On

Off

On

DSW1-7

Off

Off

On

On

Data

DSW1-8

On = 7 bit

Off = 8 bit

9600

4800

2400

1200

Baud rate

DSW1-9

Off

Off

On

On

DSW1-10

Off

On

Off

On

Computer (RS232) commands

A communication program (i.e., such as Window Terminal or ProComm
Plus) is recommended for computer operation of the instrument.
Configure the communication program to the RS232 connection settings.
Refer to

Configure the RS232 connection

on page 30.

Table 5

shows the RS232 command set for the instrument.

Table 5 RS232 command set

Command Description

VAL

Gets the current measurement with the measurement units.

LST

Gets the calibration standards and coefficients.

DAT

Gets the current date.
To change the date, enter DAT=MM/DD/YY (MM=month, DD=day,
YY=year), then push Enter.

TIM

Gets the current time in 24-hour format.
To change the time, enter TIM=HH:MM (HH=hour, MM=minutes),
then push Enter.

RMN

Gets the recorder minimum value.
To change the recorder minimum value, enter RMN=XXXXX
(XXXXX=a number, minimum value=0), then push Enter.

RMX

Gets the recorder maximum value.
To change the recorder maximum value, enter RMX=XXXXX
(XXXXX=a number, maximum value=10,000), then push Enter.

RTN

Gets the recorder trim minimum value.
To change the recorder minimum value, enter RTN=XXXXX
(XXXXX=a number, minimum value=200), then push Enter.

RTX

Gets the recorder trim maximum value.
To change the recorder maximum value, enter RTX=XXXX
(XXXX=a number, maximum value=4800), then push Enter.

SAV

Gets the signal average buffer size.
To change the signal average buffer size, enter SAV=XX (XX=a
number, maximum value=15, default=10), then push Enter.

English 31