Computer (rs232) commands, Figure 10, Table 4 – Hach-Lange 2100 N Laboratory Turbidimeter User Manual User Manual
Page 33
1. Set switches 1, 2, 3, 4, 5 and 8 to OFF.
2. Set switches 6, 7, 9 and 10 to ON. Switch 9 and 10 set to ON sets a
1200 baud rate.
3. Refer to Setting of Preset Jumper in the Citizen Printer Manual.
Figure 10 Citizen printer switch configuration
Table 4 Dip switch settings for the Citizen printer
USA
Germany France
UK
Foreign character selection
DSW1-4
Off
Off
On
On
DSW1-5
Off
On
Off
On
Odd
Even
No Check
Parity check
DSW1-6
Off
On
Off
On
DSW1-7
Off
Off
On
On
Data
DSW1-8
On = 7 bit
Off = 8 bit
9600
4800
2400
1200
Baud rate
DSW1-9
Off
Off
On
On
DSW1-10
Off
On
Off
On
Computer (RS232) commands
A communication program (i.e., such as Window Terminal or ProComm
Plus) is recommended for computer operation of the instrument.
Configure the communication program to the RS232 connection settings.
Refer to
Configure the RS232 connection
on page 30.
shows the RS232 command set for the instrument.
Table 5 RS232 command set
Command Description
VAL
Gets the current measurement with the measurement units.
LST
Gets the calibration standards and coefficients.
DAT
Gets the current date.
To change the date, enter DAT=MM/DD/YY (MM=month, DD=day,
YY=year), then push Enter.
TIM
Gets the current time in 24-hour format.
To change the time, enter TIM=HH:MM (HH=hour, MM=minutes),
then push Enter.
RMN
Gets the recorder minimum value.
To change the recorder minimum value, enter RMN=XXXXX
(XXXXX=a number, minimum value=0), then push Enter.
RMX
Gets the recorder maximum value.
To change the recorder maximum value, enter RMX=XXXXX
(XXXXX=a number, maximum value=10,000), then push Enter.
RTN
Gets the recorder trim minimum value.
To change the recorder minimum value, enter RTN=XXXXX
(XXXXX=a number, minimum value=200), then push Enter.
RTX
Gets the recorder trim maximum value.
To change the recorder maximum value, enter RTX=XXXX
(XXXX=a number, maximum value=4800), then push Enter.
SAV
Gets the signal average buffer size.
To change the signal average buffer size, enter SAV=XX (XX=a
number, maximum value=15, default=10), then push Enter.
English 31