beautypg.com

Specifications – Hach-Lange 2100P - Portable Turbidimeter Instrument User Manual

Page 9

background image

Page 9

4650098 specs.fm

SPECIFICATIONS

SPECIFICATIONS

Specifications are subject to change without notice. Operating specifications
applicable at 25 °C unless noted. Program software copyrighted by Hach Company,
1991.

Measurement Method: Ratio Nephelometric signal (90°) scatter light ratio to
transmitted light

Range: 0–1000 NTU (=FNU) with automatic decimal point placement or manual
range selection of 0–9.99, 0–99.9 and 0–1000 NTU (=FNU)

Accuracy: ± 2% of reading plus stray light from 0–1000 NTU

Resolution: 0.01 NTU on lowest range

Repeatability: ± 1% of reading or 0.01 NTU, whichever is greater

Response Time: 6 seconds for full step change without signal averaging in constant
reading mode

Stray Light: <0.04 NTU

Standardization: StablCal® Stabilized Formazin primary standards or Formazin
primary standards

Display: Four-digit liquid crystal; 10.16 mm high digits with custom icons

Light Source: Tungsten filament lamp

Detectors: Silicon photovoltaic

Signal Averaging: Operator selectable on or off

Sample Cells: (Height X width) 60.0 X 25 mm Borosilicate glass with screw caps,
marking band and fill line

Sample Required: 15 mL