OPERATION
Section 2 OPERATION
2.1 Operational Controls and Indicators
Key
Description
Power key turns instrument on and off. If no keys are pressed for 5.5 minutes, the instru-
ment turns off automatically.
Performs a measurement. To conserve battery power, the lamp turns on only when READ
is pressed. A reading is displayed about 12 seconds after the key is pressed. After the rea-
ding is displayed, the lamp turns off and the reading continues to be displayed. After a
measurement is taken, there is a 4-second recovery time before another can be started. If
Read is pressed during the recovery time, the display will flash but the lamp will not turn
on until the recovery time is complete. If no key strokes occur for 5.5 minutes, the meter
turns off.
Hold this key for continuous reading if the meter is not in the Signal Averaging mode.
After the initial delay, the reading is updated every 1.2 seconds.
Performs a calibration or review calibration data. Also terminates a calibration or calibra-
tion review and returns to the 2100P measurement mode.
Edits a flashing digit in the calibration mode, or scrolls through the calibration standards
(S0,S1, S2, S3) or diagnostic menu.
Moves the editing cursor to the digits being edited in the calibration mode or initiate
editing of a standard value.
Turns the signal averaging function on or off.
Selects the diagnostic mode.
2100P TURBIDIMETER
EDIT
DIAG
NTU
CAL?
POWER
MODE
CAL
DIAG
SIGNAL
AVERAGE
AUTO RNG
SIG AVG
RANGE
READ
I
O
READ
CAL
SIGNAL
AVERAGE
DIAG