4 check entered data, 5 overwrite stored calibration data, 1 retrieve the factory calibration – Hach-Lange PCII-COD LCK 014_114_314_614 User Manual
Page 30: 1 retrieve the factory, Rst. see
System start up
28
3.4 Check entered data
To avoid erroneous measurements it is recommended to check the data after
completing the programme procedure.
1.
Press MENU for about five seconds until "USER" followed by "CAL" is
displayed.
2.
Press BLUE key to scroll to "Edit". Press GREEN key.
3.
"RES" is displayed. Press BLUE key to scroll through the entries. Compare the
entries with the figures in the
.
In case a value was not stored correctly, retrieve the factory calibration first. See
Retrieve the factory calibration on page 28
. After retrieving the factory calibration,
start the programming procedure again.
3.5 Overwrite stored calibration data
With the POCKET COLORIMETER II it is possible to overwrite stored calibration
data. Before overwriting calibration data it is recommended to retrieve the factory
calibration first. After retrieving the factory calibration, you can program new
calibration data.
In case of any mistake retrieve the factory calibration first.
3.5.1 Retrieve the factory calibration
1.
Press MENU. "SEL" is displayed. A flashing arrow indicates the current
measuring range (1 or 2). Press GREEN key to toggle between measuring
ranges. Select measuring range retrieving the factory calibration.
1.
Press MENU for about five seconds until "USER" followed by "CAL" is
displayed.
2.
Press BLUE key to scroll to "dFL". Press GREEN key.