3 programming procedure, 1 lck 014 cod, 1 cod (high range lck 014/114) with – Hach-Lange PCII-COD LCK 014_114_314_614 User Manual
Page 25: Pocket colorimeter ii 600 nm
System start up
23
3.3 Programming procedure
Note: Each programming procedure has only to be followed once prior to the first test
evaluation.
Note: After programming the calibration data, it could be helpful to mark the channel(s) 1
and/or 2 accordingly with a label.
3.3.1 COD (high range LCK 014/114) with
POCKET COLORIMETER II 600 nm
3.3.1.1
LCK 014 COD
1.
Press POWER to switch on the instrument.
2.
Press MENU. "SEL" is displayed. A flashing arrow indicates the current
measuring range (1 or 2).
3.
Select measuring range 1 or 2 for LCK 014 (1000-10000 mg/l COD).
Note: Press
GREEN
key to toggle between measuring ranges.
4.
Press MENU for about five seconds until "USER" followed by "CAL" is
displayed.
5.
Press BLUE key until "Edit" is displayed. Press GREEN key. " RES" is
displayed.
6.
Press GREEN key. "-.---" is displayed.
7.
Press BLUE key until "----" is displayed. Press GREEN key. "S0" is displayed.
8.
Press GREEN key until "A0" is displayed.
9.
Press GREEN key until "S1" is displayed.
10. Press GREEN key. Enter "9999" by using BLUE key to change value and
GREEN key to move to the next space. Press GREEN key until "A1" is
displayed.
11. Press GREEN key. Enter "1.650" by using BLUE key to change value and
GREEN key to move to the next space. Press GREEN key until "Add" is
displayed.
12. Press MENU until "Edit" is displayed. Press GREEN key until „S0“ is displayed.
13. Press GREEN key. Enter "1000" by using BLUE key to change value and
GREEN key to move to the next space. Press GREEN key until "A0" is
displayed.
14. Press GREEN key. Enter "0.177" by using BLUE key to change value and
GREEN key to move to the next space. Press GREEN key until „S1“ is
displayed.
15. Press MENU until "Edit" is displayed to check the entered data. Press GREEN
key. „RES“ is displayed.