4 enter variables, 5 save a free programming user program, 3 select a user program – Hach-Lange DR 5000 User Manual User Manual
Page 82: 4 enter variables 6.1.2.5 save a free, 3 select, Advanced operations

82
Advanced Operations
6.1.2.4 Enter variables
1. Highlight the Variables line in the data overview and press
Edit.
2. Select the variable to be edited, press Edit and use the
alphanumeric keypad to enter the data specified in the working
procedure (for F1, F2,
λ1, U1 etc.). Press OK to confirm each
entry.
Abbreviation of Variables:
F1: Factor 1
F2: Factor 2
λ1: Wavelength 1
U1: Conversion Factor 1 for the first chemical form
U2: Conversion Factor 2 for the further chemical form etc.
6.1.2.5 Save a free programming user program
1. Press Store to save the entered data. The data can be stored
under any data point (Measurement sequence, Formula, Timer,
etc.).
2. Press the PC & Printer icon to send the program data to a
printer or to a USB memory stick (connect the USB memory
stick to the USB interface first).
Note: The test data will be formatted in the preinstalled PrgData folder
as a .csv and as a .lst file.
3. To transfer the program data from one instrument to another
create a new folder on the USB memory stick named dbhl and
copy the .lst files into this folder (see
).
6.1.3 Select a user program
1. Press User Programs in the Main Menu to view an
alphabetical list of user programs with program numbers.
The User Programs list will appear.
Note: Use the scroll bar to scroll through the list quickly.
Note: Highlight the selection by pressing it or press Select by
Number to search for the program by number.
2. Press Start to run the program.