4 matx2 cl- correction, Tx2 cl, Operation – Hach-Lange NO3D sc Nitrate Sensor User Manual
Page 26

26
Operation
After determining the laboratory reference value, proceed as
follows:
7. Select
SENSOR SETUP>NO3D SC>CALIBRATE>
SET NO3–N CONC
.
8. Enter the laboratory value (reference value) for NO
3
–N and
press
ENTER
to confirm.
9. Select
MEAS CONC 2
and repeat the sequence from Point 4 to
8 for the second value after a concentration change of at least
half a decade.
10. By confirming the second laboratory value, the Matrix2
correction is activated.
4.6.3.4 MATX2 CL- correction
Select
MATX2 CL-
to calibrate the nitrate and chloride sensor
simultaneously at two different measuring points.
1. Select
SENSOR SETUP>NO3D SC>CALIBRATE> CAL CONFIG
.
2. Select
MATX2 CL-
in the selection window and press
ENTER
.
3. Select
CONC MEAS 1
and wait until the NO
3
–N and Cl
-
values
are stable (drift should be <0.03 mg/L).
4. Press
ENTER
to confirm. The value for nitrate and chloride are
saved.
5. Immediately after measuring, take a water sample for
laboratory analysis from as close to the sensor as possible.
6. Analyze the sample immediately after taking the sample as its
nitrate content can change quickly.
After determining the laboratory reference value, proceed as
follows:
1. Select
SENSOR SETUP>NO3D SC>CALIBRATE>SET NO3-N
CONC
.
2. Enter the laboratory value (reference value) for NO
3
-N and Cl
-
press
ENTER
to confirm.
3. Select
MEAS CONC 2
and repeat the sequence from
MEAS
CONC 1
for the second value after a concentration change of at
least half a decade.
4. By confirming the second laboratory value, the Matrix2
correction is activated.
CALIBRATE
CAL. CONFIG.
CONC MEAS 1
DATE
SET CL- CONC
SET NO3–N CONC
CONC MEAS 2
DATE
SET CL- CONC
SET NO3–N CONC