5 sensor setup menu, Operation – Hach-Lange NO3D sc Nitrate Sensor User Manual
Page 20

20
Operation
4.5 Sensor setup menu
SELECT SENSOR (if there is more than one sensor)
CALIBRATE
CAL.CONFIG.
Note: If once a calibration method is chosen the entries will be displayed in the first
submenu of the calibrate menu.
Select SENSOR CODE, MATX1, MATX1 CL-, MATX2, MATX2 CL-, VALUE CORR,
PREVIOUS CAL or FACTORY CAL
-or-
CAL.CONFIG.>SENSOR CODE
DATE
Displays the date of sensor cartridge start up
SENSOR CODE
Display and entry of the sensor code
-or-
CAL.CONFIG.>MATX1
1 point matrix correction (see
DATE
Displays the date of the current correction
CONC MEAS 1
Saving the currently measured measurement
SET NO3–N CONC
Entry and display of the NO
3
–N reference value (lab value)
-or-
CAL.CONFIG.>MATX1 CL-
1 point matrix correction with chloride (see
)
DATE
Displays the date of the current correction
CONC MEAS 1
Saving the currently measured measurement
SET CL- CONC
Entry and display of the Cl
-
reference value (lab value)
SET NO3–N CONC
Entry and display of the NO
3
–N reference value (lab value)
-or-
CAL.CONFIG.>MATX2
2 point matrix correction (see
CONC MEAS 1
Saving the currently measured measurement
DATE
Displays the date of the current correction
SET NO3–N CONC
Entry and display of the NO
3
–N reference value (lab value)
CONC MEAS 2
Saving the second measured measurement
DATE
Displays the date of the current correction
SET NO3–N CONC
Entry and display of the NO
3
–N reference value (lab value)
-or-
CAL.CONFIG.>MATX2 CL-
2 point matrix correction with chloride (see
)
Note: Carry out the matrix correction only as a standard addition in the sample matrix.
CONC MEAS 1
Saving the currently measured measurement
DATE
Displays the date of the current correction
SET CL- CONC
Entry and display of the Cl
-
reference value (lab value)
SET NO3–N CONC
Entry and display of the NO
3
–N reference value (lab value)
CONC MEAS 2
Saving the second measured measurement
DATE
Displays the date of the current correction
SET CL- CONC
Entry and display of the Cl
-
reference value (lab value)
SET NO3–N CONC
Entry and display of the NO
3
–N reference value (lab value)