Replacement parts and accessories, Standard solutions – Hach-Lange SENSION+, Portable Meter, LPG2500 User Manual
Page 15

Error/Warning
Description
Solution
E4
Salinity cannot be
calculated TC=0
Modify TC
E5
Same buffer
solutions.
• Clean the probe and the membrane.
• Make sure that there are no air
bubbles in the membrane.
• Disconnect the probe and then
connect the probe again.
• Connect a different probe to verify if
problem is with probe or meter.
• Make sure that the buffer used
matches the buffer specified in
configuration.
• Replace the buffer solution.
• Make sure of the temperature
specification in configuration.
E6
The calibration
solutions have
different
temperatures.
Make sure that the calibration solutions
have the same temperature.
ErA (only
MM156)
Salinity is set to
automatic and
measured value >
50 g/L.
Modify TC
Replacement parts and accessories
Note: Product and Article numbers may vary for some selling regions. Contact the
appropriate distributor or refer to the company website for contact information.
Replacement parts
Description
Item no.
Batteries, Alkaline AA
1938004
Carrying case
LZW9990.99
Accessories
Description
Item no.
Electrolyte for DO probe 51 20, 25 mL
LZW9811.99
Electrolyte for DO probe 51 30, 50 ml
2759126
Tube for DO probe calibration
LZW5123.99
Service kit for 51 30 DO probe (contains 2 membrane
modules and DO filling solution)
5196800
Replacement DO membrane for DO probe 51 20
LZW5125.99
147 μS/cm standard solution (at 25 °C, 77 °F), 125 mL
bottle
LZW9701.99
1413 μS/cm standard solution (at 25 °C, 77 °F), 125 mL
bottle
LZW9711.99
12.88 mS/cm standard solution (at 25 °C, 77 °F), 125 mL
bottle
LZW9721.99
Set of calibration tubes, 147 μS/cm, 1413 μS/cm and
12.88 mS/cm
LZW9138.99
220 mV standard solution, 125 mL
LZW9402.99
Buffer solution pH 4.01, 125 mL
LZW9460.99
Buffer solution pH 7.00, 125 mL
LZW9461.97
Buffer solution pH 10.01, 125 mL
LZW9470.99
Electrolyte solution, KCI 3M, 50 mL
LZW9509.99
Calibration tubes pH 4.01, pH 7.00 and pH 10.01
LZW9137.97
Tube for ORP standard calibration 220 mV
LZW9136.99
Work protector for electrodes 50 51 T and 50 52 T
LZW9162.99
Standard solutions
Technical buffer solutions (DIN 19267)
Refer to
pH and ORP (mV) values of specific buffer sets at
varying temperatures.
English 15