Table 1 Spreadsheet column descriptions (continued)

Column header name

Data description and example values

Check Std Value

Value of the check standard that was used to verify
accuracy, for example: 7.00 pH–25 ºC (pH, temp-
compensated); 7.01 pH (pH, custom)

Check Std Units

Check standard units, for example μS/cm.

Note: pH is not displayed here as it is included in the previous
column.

Check Std Graph

Bar-graph showing the measurement in relation to the
acceptance limits. Example: “6.901 <—|—> 7.101”.

Check Std Status

Status of the check standard reading. Example:
“Reading within limits”, “Reading outside limits”

Calibration Status

Status of the calibration that is in use.

= current

calibration is valid;

= calibration has expired.

Cal Date

Date of calibration reading: stored in user-defined date
format

Cal Time

Time of calibration reading: stored in user-defined
time format

Cal Operator ID

The operator ID specified when the probe was
calibrated. Shows “- - -” if undefined.

Cal Slope Name

Slope (pH or LDO) or cell constant (conductivity)

Cal Slope

The slope value for the calibration

Cal Slope Aux

Used by pH to give the percent of nominal slope

Cal Slope Units

Units of the calibration slope. Example: “mV/pH” for
pH

Cal Offset

Calibration offset value

Cal Offset Units

Calibration offset units. Example: “mV” for pH.

Cal r²

Calibration correlation coefficient without a unit (may
be blank)

Table 1 Spreadsheet column descriptions (continued)

Column header name

Data description and example values

Cal Number of Std’s

Number of standards used during calibration, for
example 5. May be blank depending on record type,
parameter type, and method settings.

Cal Std 1

Known value of the first calibration standard

Cal Std 1 Units

Units of the first calibration standard

Cal Std 1 Primary Value Measured value of the first calibration standard

Cal Std 1 Primary Units

Associated units for the calibration measurement

Cal Std 1 Supp Value

Value of supplemental measurement, for example
temperature

Cal Std 2–7

Known value of additional calibration standards, if
used

Cal Std 2–7 Units

Units of additional calibration standards, if used

Cal Std 2–7 Primary
Value

Measured values of additional calibration standards, if
used

Cal Std 2–7 Primary
Units

Associated units for additional calibration
measurements, if used

Cal Std 2–7 Supp Value Value of supplemental measurement, for example

temperature

Cal Std Supp Units

Units applicable to all secondary calibration readings.
Example: “ºC” or “ºF” for temperature

Cal Message 1–4

Any messages about the calibration

Date/Time POSIX

Date and time of reading stored in POSIX format
(number of seconds from January 1, 1970) Example:
1149234913

Cal Date/Time POSIX

Date and time of calibration stored in POSIX format
(number of seconds from January 1, 1970). Example:
1111320348

Meter SN

Meter serial number used to take the measurement