Product overview, Calibration – Hach-Lange PHC10130 User Manual User Manual
Page 2
Specifications
Details
Dimensions (rugged)
Diameter: 46 mm (1.81 in.)
Length: 223 mm (8.73 in.)
Cable length: 5, 10, 15 or 30 m (16.40, 32.81, 49.21 or 98.42 ft)
Cable connection
M12 digital output and connector compatible with HQd meters
Product overview
The CDC401 series probe is a graphite, 4-pole conductivity probe (
). The
CDC40101 or CDC40103 standard conductivity probe is available with a 1 or 3 m (3.28 or
9.84 ft) cable and is intended for laboratory use. The CDC40105, CDC40110,
CDC40115 or CDC40130 rugged conductivity probe is available with a 5, 10, 15 or 30 m
(16.40, 32.81, 49.21 or 98.42 ft) cable. The probe measures electrical conductivity,
salinity, resistivity or total dissolved solids (TDS) in wastewater, drinking water and
general applications.
Figure 1 Probe overview
1 Temperature sensor and 4-pole graphite design
conductivity cell
3 Shroud (rugged model)
2 Standard probe (1 or 3 meter cable)
4 Rugged probe (5, 10, 15, or 30 meter cable)
Calibration
Before calibration:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
It is not necessary to recalibrate when moving a calibrated probe from one HQd meter to another if the additional meter is
configured to use the same calibration options.
To view the current calibration, push
, select View Probe Data, then select View Current Calibration.
If any two probes are connected, push the UP or DOWN arrow to change to the single display mode in order to show the
Calibrate option.
If a rugged probe, remove the shroud from the probe (refer to
Calibration notes:
•
Do not touch the tip of the probe.
•
Additional conductivity standards can be selected in the Calibration Options menu.
•
The cell constant is derived from the calibration standard.
•
Do not dilute conductivity standards and samples.
2