Advanced operation, Change measurement options – Hach-Lange ISENH318103 User Manual User Manual
Page 7

Advanced operation
Parameter-specific settings can be changed through the Full Access Options menu.
Details about menu navigation, available options and how to change them are given in
the screens, tables and procedures throughout this section.
The settings that can be changed are shown in
Table 1 Parameter-specific settings
Setting
Options
Measurement Options
•
Units
•
Significant digits
•
Auto stabilization
•
Stability criteria
•
Upper and lower range limits
Calibration Options
•
Standard set
•
Calibration units
•
Minimum calibration points
•
Slope limit
•
Calibration reminder
Check Standard Options
•
Standard
•
Check standard reminder
•
Acceptance criteria
Change measurement options
Methods are groups of factory-set or user-defined settings relevant to specific
applications. If the meter is set to a factory-set method and the Modify Current Settings
option is chosen, a prompt for a new name is shown after the changes are entered. The
settings are saved with this name to distinguish them from the factory-set methods, which
cannot be changed. A saved method can be used instead of multiple adjustments to the
individual settings. Changes made to a user-defined method are automatically saved with
the existing name. Multiple methods can be saved for the same probe on each meter.
1. Make sure a probe is connected to the meter.
2. Push and select ISENH3181 Settings.
3. Select Modify Current Settings.
4. Select Measurement Options and update the settings:
Option
Description
Chemical Form
Sets the concentration value—NH
3
or NH
3
-N.
Units
Sets the preferred unit for ISE measurements—mg/L (default), µg/L, g/L,
g/kg, mol/L, mmol/L, mol/kg, %, ppm or ppb.
Note: The mV units are shown when the detailed display is selected.
Significant Digits Sets the significant digits shown—2, 3 (default) or 4.
7