Troubleshooting, If a measurement error occurs, refer to – Hach-Lange ISENH318103 User Manual User Manual
Page 11

Note: After long-term storage, the ISENH3181 probe (with membrane module assembled)
might need to be conditioned in Ammonia probe storage solution for up to 30 minutes to
improve the stabilization speed.
Troubleshooting
Message or symptom
Possible cause
Action
Probe not supported
Software not updated
To download the most current version of the
software, refer to the applicable product page on the
manufacturer's website.
Refer to the HQd Series meter manual for specific
instructions for the meter model.
HQd meter does not support
IntelliCAL
®
probe
Contact a Technical Support Representative.
Connect a probe or probe
requires service
Probe not connected properly
Disconnect, then connect the probe. Tighten the
locking nut.
Software not updated
To download the most current version of the
software, refer to the applicable product page on the
manufacturer's website.
Refer to the HQd Series meter manual.
Large number of methods stored on
probe.
Continue to let probe connect. Do not disconnect
probe.
Damaged probe
Make sure connectivity with another probe or meter
to confirm isolated issue with probe. Contact a
Technical Support Representative.
mV reading is the same for all
solutions
Soaker bottle not removed
Remove the soaker bottle.
Electrical issue
Contact a Technical Support Representative.
Slow response time
Dirty sensing element
Clean the probe (body, membrane module and glass
bulb). Refer to
on page 10.
Membrane failure
Replace the membrane module.
Dirty filling solution
Replace the filling solution.
Low sample temperature or
temperature difference between
samples
Check the sample temperature. The lower the
temperature or the greater the difference of
temperatures between samples, traditionally the
longer the response time.
Bubbles trapped under sensor tip
Gently shake the probe until bubbles are removed
from under sensor tip.
Slope out of range (refer to
on page 13)
pH is incorrect
Make sure the pH is > 11 after each ISA addition.
Ionic strength adjustor (ISA) not
used
Add ISA to each sample and standard (one powder
pillow per 25 mL of solution).
Insufficient conditioning
Condition for at least in ammonia probe storage
solution.
Damaged probe
Contact a Technical Support Representative.
Incorrect standards
Calibrate using freshly prepared standards.
Dirty sensing element
Clean the probe (body, membrane module and glass
bulb) and recalibrate.
Bubbles trapped under sensor tip
Gently shake the probe until bubbles are removed
from under sensor tip.
11