DOC272.53.80025
User instructions
BOD LDO
®
Probe: Model LBOD10101
Safety Information
Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed.
Overview
shows the LBOD10101 probe, a luminescent dissolved oxygen (LDO
®
)
sensor with an integrated stirring system. The LBOD10101 measures the dissolved oxygen for
BOD (Biochemical Oxygen Demand) determination in BOD bottles. The probe uses the
luminescent dissolved oxygen (LDO) technology (no membranes) and features a replaceable
self-stirring assembly and a 1.8 meter cable.
Specifications
Specifications subject to change without notice.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after
12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC),
European electrical equipment users must now return old or end-of life equipment to the Producer for disposal at no
charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return
end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
Specification
Details
% saturation resolution
0.1%
% saturation accuracy
± 0.59% of reading
Cable connection
Digital output and connector compatible with HQd meters
Dimensions
Diameter: 12 mm (0.47 in.), Length: 103 mm (4.06 in.), Total length: 220 mm (8.66 in.),
Cable length: 1.8 meter (5.91 ft)
Dissolved oxygen accuracy
± 0.05 mg/L for 0.0 to 10 mg/L
± 0.1 mg/L for greater than 10 mg/L
Dissolved oxygen range
0.05 to 20.0 mg/L (ppm) 1 to 200% saturation
Minimum sample depth
10 mm
Pressure accuracy
± 3 mbar (0.3 kPa)
Pressure resolution
5 mbar (0.5 kPa)
Response time
T90% at 10 s (when stirred)
Temperature accuracy
± 0.3 °C (± 0.54 °F)
Storage temperature
5 to 40 °C (41 to 104 °F)
Temperature resolution
0.1 °C (0.18 °F)
Temperature range
0 to 50 °C (32 to 122 °F)
Warranty
Probe and sensor cap are covered by a 1 year warranty. Use outside of the laboratory