Operation

CONFIGURE

EDIT NAME

Enter up to a 10-digit name in any combination of symbols and alpha or numeric
characters.

MEAS UNITS

Select the appropriate measurement units to display. Choose from: mg/L, ppm,
or percent. Default: ppm

TEMP UNITS

Select Celsius (°C) or Fahrenheit (°F); Default: °C

AC FREQUENCY

Choose 50 or 60 Hz depending on the power line frequency for optimal noise rejection.
Default is 60 Hz.

FILTER

Specify the number of seconds for signal averaging (0–60). Default: 0 seconds.

PRESSURE UNITS

Choose pressure units in mmHg, feet, meters. Default: mmHg

SET PRESSURE

Enter either altitude or air pressure. Correlates to the pressure units setting. Range:
–5000 to 15000. Default: 760 mmHg

SALINITY UNITS

Choose from mS/cm, mMol/L, ppt, or mg/L. Default: mS/cm

SET SALINITY

User-entered value. Default: 0.00 µS/cm

LOG SETUP

Disable or choose the datalogging interval for sensor and temperature measurements.
Use the arrow keys to move through the available choices for each. Default: Disabled

TEMP ELEMENT

Choose SELECT TYPE to specify the temperature sensor that is integrated into the
sensor or choose SET MANUAL to disable automatic temperature compensation.
Default temperature sensor: NTC 30K

CAL DAYS

Shows the number of days since the last calibration. Default reminder at 60 days.

SENSOR DAYS

Shows the number of days the sensor has been in service. Automatically reminds the
user to replace the sensor after a set period of time. Default reminder at 365 days.
Reset counter in DIAG/TEST/RESET SENSOR menu.

DEFAULT SETUP

Resets the sensor software to default settings.

DIAG/TEST

PROBE INFO

SENSOR NAME

Displays the entered name of the sensor. Default is the sensor serial number.

SERIAL NUMBER

Serial number of the sensor.

SOFTWARE VERS

Displays the software version number.

DRIVER VERS

Displays the sensor driver version number.

CAL DATA

CAL Q VALUE

Display for service diagnostics only.

OFFSET CORR

User editable—to change the calibration offset.

LAST CAL DATE

Shows the date of the last calibration. Default: 1-1-00

SIGNALS

SENSOR SIGNAL

Displays the sensor output in mV.

SENSOR ADC COUNTS

Raw data for sensor ADC counts. Comparable to A/D counts.

TEMP ADC CNTS

Raw data for temperature ADC counts. Comparable to A/D counts.

COUNTERS

SENSOR DAYS

Cumulative days the sensor has been in use.

RESET SENSOR

Resets sensor counter.

4.6 SENSOR SETUP Menu (continued)

SENSOR SETUP

Select Sensor (if more than one sensor is attached)