Programming of the transmitter, Setting the type of measurement – Hach-Lange POLYMETRON 83xx User Manual User Manual
Page 18

2 E
LECTRODES
C
ONDUCTIVITY
P
ROBE
- I
NSTRUCTION
M
ANUAL
Page 14
09/2010
621=183=000 Issue B
Programming of the transmitter
REMARK
To obtained detailed information, please
refer to the operating manuals of our
transmitters.
Setting the type of measurement:
9125: Check that both switches of the
conductivity module have been correctly
configured on K (2 electrodes).
Setting the frequency in function with the
conductivity:
K (cm
-1
)
Low
conductivity
Average
conductivity
High
conductivity
0.01
0.01…0.1 µS
0.1 µS…20 µS
20 µS…200 µS
0.1
0.1….1 µ S
1 µS…200 µS
200µS…2 mS
1
1 … 10 µ S
10 µS…2 mS
2…20 mS
It is preferable, whenever possible, to operate
in the "Average conductivity" zone (and
therefore to choose the type of sensor well).
• « Low conductivity » zone:
Do not combine a long length of cable with
a high measurement frequency, to avoid
provoking a parallel capacitance
(measurement of conductivity too high).
If one uses a long cable (> 20 m), adjust to
a frequency of 70 Hz.
- AMTAX sc (118 pages)
- PHOSPHAX sc (106 pages)
- CL17 Instruction sheet (8 pages)
- CL17 USER MANUAL (60 pages)
- CLF10 sc (382 pages)
- 9184 sc (50 pages)
- POLYMETRON 9611 sc Installation (398 pages)
- POLYMETRON 9610 sc Maintenance and Troubleshooting (286 pages)
- POLYMETRON 9611 sc USER INSTRUCTIONS: SEQUENCER LINE INSTALLATION (2 pages)
- POLYMETRON 9611 sc USER INSTRUCTIONS: SS SAMPLE CONDITIONING (4 pages)
- HACH 5500 sc Installation (408 pages)
- HACH 5500 sc Maintenance and Troubleshooting (286 pages)
- HACH 5500 sc PO43-HR Operations (390 pages)
- HACH 5500 sc PO43-LR Operations (392 pages)
- HACH 5500 sc SiO2 Installation (418 pages)
- HACH 5500 sc SiO2 Operations (394 pages)
- POLYMETRON 9610 sc SiO2 Installation (410 pages)
- POLYMETRON 9240 Basic User Manual (162 pages)
- POLYMETRON 9240 Operator Manual (144 pages)
- PHOSPHAX sigma Basic User Manual (305 pages)
- AMTAX inter2 Basic User Manual (377 pages)
- GANIMEDE P (60 pages)
- BODTrak II Basic User Manual (172 pages)
- BODTRACK II User Manual (28 pages)
- BODTRAK II: Respirometric Biochemical Oxygen Demand (BOD) (46 pages)
- SPECTRO COLOR d-8 (29 pages)
- REFO 60_REFO 60D (36 pages)
- REFO 3_REFO 3D (19 pages)
- LICO 620 Basic User Manual (329 pages)
- LICO 620 User Manual (114 pages)
- SD900 (2 pages)
- LZX971 (8 pages)
- SC Sensor Plug and SC Sensor Coupling (8 pages)
- HQ Series Portable Meters (4 pages)
- LDO Sensor Replacement Kit for use with Hach HQ10 and HQ20 (43 pages)
- LDO Sensor Replacement Kit (8 pages)
- SIP 10 (273 pages)
- SC 100 (58 pages)
- SI792 E_T Quick start guide (2 pages)
- SI792(x) E HACH GLI 3700 series SI792(x) T 7MA2200 and 8398 series User Manual (116 pages)
- SI792 D Quick start guide (2 pages)
- SI792 D, SI792x D, SI792x D-FF, SI792x D-PA User Manual (98 pages)
- SI794 C and SI794 D Short instructions (2 pages)
- SI794 D tr User Manual (108 pages)
- SI792 C Quick start guide (2 pages)