background image

2 E

LECTRODES

C

ONDUCTIVITY

P

ROBE

- I

NSTRUCTION

M

ANUAL



Page 14

09/2010

621=183=000 Issue B

Programming of the transmitter

REMARK

To obtained detailed information, please
refer to the operating manuals of our
transmitters.

Setting the type of measurement:

9125: Check that both switches of the
conductivity module have been correctly
configured on K (2 electrodes).

Setting the frequency in function with the
conductivity:

K (cm

-1

)

Low

conductivity

Average

conductivity

High

conductivity

0.01

0.01…0.1 µS

0.1 µS…20 µS

20 µS…200 µS

0.1

0.1….1 µ S

1 µS…200 µS

200µS…2 mS

1

1 … 10 µ S

10 µS…2 mS

2…20 mS


It is preferable, whenever possible, to operate
in the "Average conductivity" zone (and
therefore to choose the type of sensor well).

« Low conductivity » zone:

Do not combine a long length of cable with
a high measurement frequency, to avoid
provoking a parallel capacitance
(measurement of conductivity too high).


If one uses a long cable (> 20 m), adjust to
a frequency of 70 Hz.