Warning list – Hach-Lange POLYMETRON 9526 User Manual
Page 29
1. Push the menu key and select DIAGNOSTICS.
2. Select the device (controller, sensor, network card) with the warning or error and push enter.
3. Select the warning, error or event list and push enter.
4. Select YES and push enter to acknowledge the warning.
Note: Errors cannot be acknowledged.
Warning list
A warning icon consists of an exclamation point within a triangle. Warning icons appear on the right
of the main display below the measurement value. A warning does not affect the operation of menus,
relays and outputs. To view warnings, push the menu key and select DIAGNOSTICS. Then select
the device to view any problems associated with that device. The warning icon will no longer be
displayed once the problem has been corrected or acknowledged.
A list of possible warnings is shown in
.
Table 3 Warning list for conductivity sensors
Warning
Description
Resolution
MEAS TOO HIGH
The measured value is > 2 S/cm,
1,000,000 ppm, 200% or 20,000 ppt
Make sure that the display format is
set for the correct measurement
range
MEAS TOO LOW
The measured value is < 0 μS/cm,
0 ppm, 0% or 0 ppt
Make sure that the sensor is
configured for the correct cell
constant.
ZERO TOO HIGH
The zero calibration value is too
high
Make sure that the sensor is held in
air during zero calibration and is not
located near radio frequency or
electromagnetic interference. Make
sure that the cable is shielded by
metal conduit.
ZERO TOO LOW
The zero calibration value is too low
TEMP TOO HIGH
The measured temperature is >
200 °C
Make sure that the sensor is
configured for the correct
temperature element.
TEMP TOO LOW
The measured temperature is <
-20 °C
CAL OVERDUE
The Cal Reminder time has expired
Calibrate the sensor.
NOT CALIBRATED
The sensor has not been calibrated
Calibrate the sensor.
REPLACE SENSOR
The sensor has been in operation >
365 days
Calibrate the sensor with a
reference solution and reset the
sensor days. Refer to
on page 26. If the calibration fails,
call technical support.
CAL IN PROGRESS
A calibration was started but not
completed
Return to calibration.
OUTPUTS ON HOLD
During calibration, the outputs were
set to hold for a selected time.
The outputs will become active after
the selected time period.
WRONG LINEAR TC
The user-defined linear temperature
compensation is out of range
The value must be between 0 and
4%/°C; 0 to 200 °C.
WRONG TC TABLE
The user-defined temperature
compensation table is out of range
The temperature is above or below
the temperature range defined by
the table.
WRNG USER CONC TABLE
The concentration measurement is
outside of the range of the user
table
Make sure that the user table is set
for the correct measurement range.
English 27