Messages.message out.warnings" item, Messages.message out.message in last error" item – Hach-Lange ORBISPHERE 6110 OPC Instruction Manual User Manual
Page 14

Error value
Error description
0x00040000
Antifoam counter
0x00080000
Pending action
0x00100000
Pending manual action
0x00200000
No flow
0x00400000
RIM detector error
0x00800000
Upper limit
0x01000000
System initialization required
0x02000000
Communication error
0x04000000
ADIO board KO
0x08000000
CTRL board KO
0x10000000
EC board KO
0x20000000
CO2 foam out of specifications
0x40000000
Foam too many times
"Messages.Message Out.Warnings" item
This output item provides information about any warning(s) associated with the system. A warning
does not prevent the instrument from running a measurement analysis. This is a flag field where the
indicated value is computed by the addition of all active warnings.
The list items are identical to those listed in the corresponding errors table (see
on page 11).
"Messages.Message Out.Message In Last Error" item
This item indicates the last error raised by an incoming OPC message. For example, when a
"Message In.New Package" item (see
"Messages.Message In.New Package" item
on page 10)
requests setting the package to one that is not defined on the 6110 instrument, the error value will be
"POOR".
Note: In the following table "par_name" represents the parameter name to which the error relates.
Error value
Error description
NONE
No error
UNKN par_name
Unknown parameter
POOR par_name
Parameter out of range
TYPE par_name
Parameter wrong type
DENY par_name
Parameter access denied
NR par_name
Not ready, a command has been received at the wrong
time
12 English