Measure.warnings" item, Messages.message in.user id" item, Messages.message in.user password" item – Hach-Lange ORBISPHERE 6110 OPC Instruction Manual User Manual
Page 12: Messages.message in.new package" item

Error value
Error description
0x0080
No package
0x0100
Ultrasound
0x0200
Upper limit
0x0400
Clear flag detection
0x0800
P2 too low
0x1000
HS pressure unstable
0x2000
Abort
0x4000
Wrong computation error
0x8000
Unknown error
"Measure.Warnings" item
This item provides a list of any warning(s) associated with the measurement. This is a flag field
where the indicated value is computed by the addition of all active warnings.
Warning value
Warning description
0x0000
No warning
0x0001
Unstable reference pressure
0x0002
Package unmatched
0x0004
US time out gas exp.
0x0008
Reserved
0x0010
HS O2 unstable
0x0020
Insufficient gas
0x0040
Degas
0x0080
US insufficient gas
0x0100
Temperature time out
0x0200
HS pressure too low
0x0400
CO2 T2 out of specifications
"Messages.Message In.User ID" item
This item contains the user ID defined on the 6110 instrument.
"Messages.Message In.User Password" item
This item contains the user password defined on the 6110 instrument.
"Messages.Message In.New Package" item
This item specifies the index number of the next package to analyze. The index number starts at '1'
and the maximum value is defined by the 6110 instrument. The index number must be a valid
package number on the 6110 instrument.