3 programming procedure, 1 lck 348 phosphate-phosphorus po – Hach-Lange LCK 348_349_350 User Manual
Page 23

23
3.2.2.2
Evaluation as Phosphate PO
4
3.3 Programming procedure
Note: Each programming procedure has only to be followed once prior to the first test
evaluation.
Important Note: The following programming procedures are valid for the evaluation
form Phosphate-Phosphorus PO
4
-P. To evaluate as Phosphate PO
4
please choose
3.2.2.2 Evaluation as Phosphate PO4 on page 23
Note: After programming the calibration data, it could be helpful to mark the channel(s) 1
and/or 2 accordingly with a label.
3.3.1 LCK 348 Phosphate-Phosphorus PO
4
-P
1.
Press POWER to switch on the instrument.
2.
Press MENU. "SEL" is displayed. A flashing arrow indicates the current
measuring range (1 or 2).
3.
Select measuring range 1 or 2 for LCK 348 (0.5-5 mg/l PO
4
-P).
Note: Press
GREEN
key to toggle between measuring ranges.
4.
Press MENU for about five seconds until "USER" followed by "CAL" is
displayed.
5.
Press BLUE key until "Edit" is displayed. Press GREEN key. " RES" is
displayed.
6.
Press GREEN key. "-.---" is displayed.
7.
Press BLUE key until "--.--" is displayed. Press GREEN key. "S0" is displayed.
8.
Press GREEN key until "A0" is displayed.
9.
Press GREEN key until "S1" is displayed.
10. Press GREEN key. Enter "05.00" by using BLUE key to change value and
GREEN key to move to the next space. Press GREEN key until "A1" is
displayed.
11. Press GREEN key. Enter "0.600" by using BLUE key to change value and
GREEN key to move to the next space. Press GREEN key until "Add" is
displayed.
Table 2 Evaluation as Phosphate PO
4
LCK 348
LCK 349
LCK 350
RES (resolution)
--.--
--.--
--.--
S0 (standard 0)
01.50
00.15
06.00
A0 (absorbance 0)
0.104
0.090
0.104
S1 (standard 1)
15.00
04.50
60.00
A1 (absorbance 1)
0.600
0.728
0.753