Troubleshooting
Error messages
Potential SIP 10 errors are displayed by the photometer.
Replacement parts
Table 3 Error messages
Error displayed
Cause
Resolution
Sipper module not
connected.
Please check cable.
Connection between
SIP 10 and photometer
is disrupted.
Check the USB cable.
The cable length must not
exceed 1 m (3.3 ft) and the
cable must be directly
connected to the photometer.
Remove other devices that are
connected where necessary.
Please check sipper
and tube.
Pump tubing is not
correctly inserted.
Loosen the cover and position
again. Where required, run the
pump briefly before the lever is
moved to make sure that the
tube is correctly positioned
around the rollers.
Designation
Order no.
SIP 10 Sipper module set for DR 3900 complete with tube
set and 1 inch/1 cm Pour-Thru Cell, EU
LQV157.99.10001
SIP 10 Sipper module set for DR 6000 complete with tray,
tube set and 1 inch/1 cm Pour-Thru Cell, EU
LQV157.99.20001
SIP 10 Sipper module set DR 6000 complete with tray, tube
set and 1 cm quartz glass Pour-Thru Cell, EU
LQV157.99.30001
SIP 10 Sipper module; including pump tubing, EU
LQV157.99.00001
USB cable, type AB 1 m (3.3 ft)
LZQ104
Dual Path Length Pour-Thru Cell, plastic, 1 inch/1 cm
LZV876
UV Pour-Thru Cell, quartz glass, 1 cm
LZV510
Pump tubing (Lagoprene
®
) pre-assembled with cover and
connections
LZV877
Complete tube set for DR 3900, including Tygon tubing,
1.70 m (5.6 ft), pump tubing (Lagoprene pre-assembled with
cover and connections), stainless steel tube adapter
LZV875
Complete tube set for DR 6000 and Dual Path Length Pour-
Thru Cell, including Tygon tubing and Pharmed tubing,
pump tube (Lagoprene pre-installed with cover and
connections), stainless steel tube adapter
LZQ102
Complete tube set for DR 6000 "drinking water application,"
including Pharmed tubing, pump tube (Lagoprene pre-
installed with cover and connections), stainless steel tube
adapter
LZQ100
Tray for storage
HTT057
Cell compartment cover
LZQ105
User manual, multilingual SIP 10, EU
DOC012.98.90328
Information on Tygon tubes, EC no. 1907/2006 (REACH)
DOC063.98.90334