No valid data found!
Data cannot be viewed in the data log
Change the selection.
No measurement data present!
Data analysis settings cannot be configured without
measurement data.
Change the selection.
Control range not reached!
Data analysis limits not reached
This is a warning notice. The control limit set was not
reached.
Control range exceeded!
Data analysis limits exceeded.
This is a warning notice. The control limit was
exceeded.
Concentration too high!
Calculated concentration is higher than 999999
Dilute sample and measure again
Possible interference by:
Interference Check
The analysis is possibly erroneous due to
interferences.
Possible interference from:
Interference Check
The analysis is possibly erroneous due to
interferences.
Next service is due!
Contact the manufacturer or national agent for an
instrument service.
Negative result!
The calculated result is negative
Check concentration of sample
Network switched off.
Network setup off, when access to homepage via
sidebar
Activate the online connection.
Remote server cannot be reached.
Error during network setup
Make sure that the instrument is connected to the
network.
Unstable lighting conditions!
Avoid direct sunlight at the measuring location.
Please insert adapter Z.
For measurements with 11 mm round cuvettes, adapter
Z is required.
Insert adapter Z into cell compartment (2). Confirm with
OK.
Insufficient memory for update .
Error during update.
Select a memory with more space.
System check incorrect!
Measurement of air values failed
Switch the instrument off and then back on again. If the
system test is unsuccessful, contact the manufacturer
or the national agent.
Temperature too high.
Measurement not possible!
Switch of the instrument and allow it to cool for a few
minutes. If necessary, move it to a cooler place.