26 problem probable cause remedy – Hydrotech MP Series PULSAtron User Manual
Page 26

26
PROBLEM
PROBABLE CAUSE
REMEDY
1.
Dirty check valve
1.
Remove and replace or clean off any scale or sediment.
PUMP
2.
Ball checks not seating or
2.
Check seat and ball checks for chips, clean gently. If
LOSES
not sealing properly
deformity or deterioration is noted, replace part with
PRIME
proper material. Resulting crystals can hold check
valves open, therefore the valves must be disassembled
and cleaned. Be sure to replace all parts as shown in
the Parts Diagram (at the end of the manual).
3.
Solution container allowed
3.
Refill the tank with solution and prime. (See Start-Up
to run dry
and Operation Section).
4.
Chemical Outgassing
4.
Bleed gas, use flooded suction, maintain chemical at room
temperature (approx. 20
K
F / -6
K
C).
1.
Loose fittings
1.
Tighten hand tight. Replace gasket if hand tight does
not stop leakage
LEAKAGE
AT
FITTING
2.
Broken or twisted gasket
2.
Check gaskets and replace if broken or damaged.
3.
Chemical attack
3.
Consult your pump supplier for alternate material.
1.
Too much pressure at
1.
Turn off all pressure valves, loosen outlet tubing
discharge
connection at discharge point. Remove discharge valve
cartridge. Dampen ball check and valve seats with a few
drops of solution. Set pump dial to maximum rate.
When pump is primed, reconnect all tubing connections.
2.
Check valves not sealing
2.
Disassemble, loosen, clean and check for deterioration
PUMP WILL
swelling. Reassemble and wet the valve assembly,
NOT PRIME
then prime. See Start-Up and Operating Section.
3.
Output dials not set at
3.
Always prime pump with output dial set at maximum
maximum
rated capacity.
4.
Suction lift height too much
4.
Decrease suction lift or pull vacuum on pump discharge
until pump is primed.
5.
Pump equipped with spring
5.
Loosen discharge valve to aid in priming, take necessary
loaded high viscosity valves
safety precautions. Or apply vacuum to pump discharge.