SECTION - 4
OPERATING INSTRUCTIONS
4.1
ACCEPTANCE TEST :
These acceptance tests ensure that the unit is in working
order after shipping. These tests may also be used to
verify basic operation when the unit has been restored
to normal use after prolonged storage.
4.2
VARIABLE OUTPUT SECTIONS
4.2.1
CONSTANT VOLTAGE MODE :
a)
Ensure that the AC power switch is in the OFF position.
b)
Connect the unit to a rated power source.
c)
Turn the voltage and current controls fully counter
clockwise.
d)
Connect a digital voltmeter (DVM) to the output
terminals, observing correct polarirty. The DVM must be
rated better than 0.5% accuracy.
e)
Turn the power switch ON. The front panel digital meters
will light up and voltmeter and ammmeter displays will
read zero.
f)
Turn the current control a 1/2 turn clockwise. Slowly turn
the voltage control clockwise and observe both the front
panel voltmeter and the DVM.
g)
Compare the DVM reading with the front panel
voltmeter reading. The control range will be from 0 to
32V. The green voltage mode LED will be illuminated to
indicate CV mode of operation.
4.2.2
CONSTANT CURRENT MODE :
a)
Ensure that the AC power switch is in the OFF position.
b)
Connect the unit to a rated power source.
4
PCB Components
ZSDT-CT/05 PCB
=======================================================================
Ref Designator
Value
=======================================================================
C24
10µF/50V, ELE
C25
10µF/50V, ELE
C26
0.1µF/50V, CD
C27
470µF/50V, ELE
C29
1000µF/35V,ELE
DIODES
CR1
Not Used
CR2
1N4007, 1KV/1A
CR3
1N4007, 1KV/1A
CR4
1N4007, 1KV/1A
CR5
1N4007, 1KV/1A
CR6
1N4007, 1KV/1A
CR7
1N4007, 1KV/1A
CR8
1N4007, 1KV/1A
CR9
1N4007, 1KV/1A
C10
1N4007, 1KV/1A
CR11
1N4007, 1KV/1A
CR12
1N4007,1KV/1A.
CR13
1N4007,1KV/1A.
CR14
1N4007,1KV/1A
CR15
1N4007,1KV/1A
CR16
1N4007, 1KV/1A
CR17
1N4007, 1KV/1A
CR18
1N4007, 1KV/1A
CR19
1N4007, 1KV/1A
CR20
1N4007, 1KV/1A
CR21
1N4007, 1KV/1A
CR22
1N4007, 1KV/1A
11