Brooks Instrument Model 1307 O-ring Seal Flowmeter
1-4
Table 1-2. Pressure Ratings.
Meter
Maximum Working
Max.
Size
Pressure (psig)
Temp.
Gas
Service
7
350
8
300
9
175
250°F
10
100
(121°C)
12
75
13
100
1. Pressure ratings are based on static pressure applicable at
250°F.
2. Glass tubes are not usable for caustic service.
3. Fluid temperatures below 32°F will cause frosting of the glass
metering tube. Please consult factory for applications below
this temperature.
Figure 1-1. Float Types.
READ
HERE
Spherical -
Lowest meter capacity and
no viscous fluids with medium
capacity.
Type LJ -
Maximum flowmeter capacity
with limited viscosity
immunity.
READ
HERE
Type RV (rib guided) -
Highest immunity to viscous
fluids with medium meter
capacity. (Most stable)
Type GV (rod guided) -
Analogous in design to a
sharp edge orifice. Gives
most immunity to viscosity
variations of the fluid being
metered.
READ
HERE
Type RV
(Rib guided)
Type RS (rib guided) - High
flow capacity with some
immunity to viscous fluids.
Type GS (rod guided) -
Analogous in design to a flow
nozzle or venturi. Offers the
maximum flow capacity in any
given size flowmeter.
READ
HERE
Type RS
(Rib guided)
Type LJ
Spherical