SPECIFICATIONS
Step Two
Service:
Compatible liquids
Wetted Materials:
Body: 316 SS, 316L SS and
Buna‐N
Cable: Polyurethane or ETFE
Bullet Nose: PVC (LD31
series only)
Accuracy:
±0.25% of full scale
Temperature Limit:
0 to 176°F (‐18 to 80°C)
Compensated
Temperature Range:
0 to 176°F (‐18 to 80°C)
Thermal Effect:
Less than ±0.02%/°F
Pressure Limit:
2X full scale
Power Requirement:
10 to 28 VDC
Output Signal:
4 to 20 mA DC, 2‐wire
Response Time:
50 ms
Max. Loop Resistance:
900 ohms
Electrical Connections:
Wire pigtail
Cable Length:
‐__01/‐__11: 40’ (12.2 m)
‐__21/‐__31: 60’ (18.3 m)
Mounting Orientation:
Suspended in tank below
level being measured ‐ Can
be placed on the bottom of
the tank on its side
Weight:
2.2 lb (1.0 kg) – (LD31
series)
4.3 lb (2.0 kg) – (LD32
series)
Agency Approvals:
CE, UL Intrinsically Safe to
UL Standard 913. (See
Intrinsic Safety Approval
Classification.)
The following standards were used for CE approval:
IEC 61000‐4‐2: 2001
IEC 61000‐4‐3: 2006
IEC 61000‐4‐4: 2004
IEC 61000‐4‐5: 2005
IEC 61000‐4‐6: 2006
IEC 61000‐4‐8: 2001
CENELEC EN 55011: 2003
CENELEC EN 61326: 2003
89/336/EEC EMC Directive
WARNING: To prevent ignition of flammable or
combustible atmospheres, disconnect power before
servicing. Use with approved safety barriers using
entity evaluation.
Series
LD31
LD32
Vmax
26 VDC
28 VDC
Imax
93 mA
93 mA
Ci
0.051uF
0.051 uF
Li
240 uH
240 uH
Pi
0.651 W
0.651 W
WARNING: Do not exceed specified supply voltage
ratings. Permanent damage not covered by warranty
will result. This device is not designed for 120 or 240
volt AC operations. Use only on 10 to 28 VDC.
Optional Desiccant Filter (LD90‐3000):
o
For extra protection against humidity we offer the LD90‐3000 desiccant filter that can be attached to
the vent tube.
Optional Cable Hanger (LD90‐S000):
o
Attach the cable hanger to the ½” thread of the LD31 or LD32 series and
connect to the loop a chain to provide a secure method for lowering and
raising the sensor.
o
The cable hanger is ideal for sumps or wet wells.