The following list outlines how to install the DT6
transducer for optimal performance, highest reliability
and greatest accuracy:
1. Select a transducer site at least 10 pipe diameters
downstream from bends, or fittings and 5 pipe
diameters upstream. A symmetrical flow pattern is
necessary for accuracy and repeatability over the
PART 2 - TRANSDUCER INSTALLATION
Table 2.1
1
1
The D601 system will provide repeatable measurements on piping systems that do not meet these
requirements, but the accuracy of these readings may be influenced to various degrees.