Part 3 - trouble shooting – Dynasonics D601 Series Doppler Ultrasonic Flow Meter User Manual
Page 34

Rev. 4/00 -3.21- D601
PART 3 - TROUBLE SHOOTING
 
Stability of flow readings are 
unsatisfactory 
•
Increase the DAMP constant from keypad.
•
Move transducers to a location further 
from piping tees, elbows, valves, filters, 
 
Erroneous Reading 
•
Transducer mounted incorrectly or not 
true to the pipe. 
•
Another local ultrasonic instrument is 
operating at about the same frequency 
[ consult the Dynasonics factory]. 
•
Presence of large amounts of suspended 
solids or aeration. Use CAL constant to 
compensate. 
•
Sources of radiated interference are 
present. Apply appropriate shielding. 
•
An electrically noisy power supply is 
powering the D601. Power the meter with 
a circuit that does not power motors, 
 
The D601 display indicates flow, when 
true fluid velocity is zero. 
•
Verify that residual leakage and flow is not 
present. [I.e. leaking check valves] 
•
Verify that GAIN control (R13 on the Main 
PCB) is not adjusted too high. With 
nominal flow running through the pipe, 
adjust GAIN control till the display zeros 
with no flow. 