1.
Connect to isolated battery terminals only
2.
Make good connection RED +, BLACK -
3.
All LEDs should indicate in sequence before commencing test
4.
Provided power is available in battery under test, a flashing YELLOW LED will indicate
battery load test in progress
5.
Latched RED LED indicates battery under test has failed
6.
Latched GREEN LED indicates battery under test has passed
7.
If required, test twice in succession only to simulate a 3 hour load test, leaving 30%
minimum capacity available in the battery
NOTE: When flashing RED LED indicates, allow battery tester 5 minutes to cool down
IMPORTANT: Only use test leads provided (Part: CD4000TL) Replace polarity protection fuse
with 10Amp blade type only
CD4000
Luminaire Battery Tester
Simulates a 1 hour battery load test in under 6 minutes
Tests all 2.4 - 12V rechargeable batteries above 1.5AH
FAIL
TESTING
PASS
Operating Instructions
CD4000
Luminaire Battery Tester
Take minutes not hours to test luminaire batteries used in stand-alone
emergency lighting systems. The CD4000 performs 1 and 3 hour duration
tests in minutes on 2.4V - 12V rechargeable Lead-acid, Lithium-ion, Nickel-
cadmium and Nickel Metal Hydride batteries.
Meters Ltd
www.actmeters.com
+44(0)1744 886660
Input Sockets
RED + BLACK -
Test Leads
2m long, heavy duty leads with crock clips
(ACT CD4000TL)
Power Supply
Self powered by the battery under test
(step-up inverter)
LED Indication
RED/Fail YELLOW/Testing GREEN/Pass
Acceptable Batteries
2.4V to 12V SLA, LI-ION, NICAD & NIMH
Minimum Battery Capacity
Greater than 1.5Ah
Current Drawn During Test
0.5Amp
Test Times
Normally under 6mins dependant on
battery voltage
Heat Dissipation
Between 23 to 115 watts
Ambient Operating Temperature
-5 to 40C
Maximum Heatsink Temperature
(Safety Trip Point)
100C
Input Protection
Reverse polarity and high voltage protection
(18VDC Max)
Input Fuse
10Amp quick blow blade type
Enclosure
Flame retardant plastic
Size
167mm x 107mm x 65mm
Weight
1.5kg
Specifications
2004/108/EC
CALINST/0213
ACT Meters Ltd
The Old Smithy
Church Road
Rainford
Merseyside
WA11 8HD
United Kingdom
w: www.actmeters.com
t: +44(0)1744 886660
f: +44(0)1744 886661