Introduction

The C-Squared Oxygen Analyzer is designed to measure oxygen levels in the 0.1 to 100.0%
range. It can be used to measure the oxygen content for mixes used in Diving, Medical
Instrumentation and Industrial applications. The Analyzer is designed to be able to verify
oxygen concentration in cylinders as well as analyze delivery systems.

The C-Squared Analyzer has a digital display, an internally or externally mounted temperature
compensated sensor with a 2 year life expectancy. The power is supplied by three AAA 1.5
volt batteries.
The case has an environmentally sealed, potentiometer, on/off/on with back light switch, long
life sensor and digital panel meter. The Analyzer is water resistant and sealed to prevent
moisture from affecting the internal components.

The Analyzer is shipped in a high impact case that can be used for storage or other purposes.
The Analyzer is ready to use out of the box with no additional work other than checking on the
calibration either in air (20.9%), or 100% O2.

Controls

The C-Squared Oxygen Analyzer has a splash proof on/off/on (with back light). When the
switch is centered it is in the off position. When not in use, the switch should be in the off
position to conserve battery life. When the switch is in the up position it is on. When the
switch is flipped down, the switch is in the on position with the panel meter back light activated.
The back light option should only be used in limited light conditions to extend the life of the
batteries.

When switched on, the analyzer’s display will indicate an oxygen reading. Please note that the
analyzer must be calibrated prior to use (see calibration section).

The low battery warning is a battery symbol in the left corner of the display. When present
change the batteries prior to using the analyzer.

The display has a 0.1% resolution and is sealed. It also can be backlit (controlled via the
on/off switch).

Incorporated

Oxygen Analyzer

Instruction Manual