Specifications, Technical data, 2 service information – Carolina Electronic Balances User Manual
Page 11: 1 specifications, Batt loww, Table 6-1.specifications
Symptom Possible
Cause
Remedy
Err 2
• Over-load or Under-load
condition.
• Transportation Lock is set
• Improper calibration
• Verify pan is installed and/or excess load is
removed from pan.
• Slide the Lock pin to the unlocked position.
• Perform calibration.
Err 4
• Invalid calibration mass.
• Verify correct calibration mass is being used.
Err 8
• RS232 buffer is full.
• Cycle scale Off/On.
• Check RS232 settings in scale and computer
when RS232 option is installed.
Err 9
• Internal data error.
• Contact an authorized dealer.
(Flashing)
• Battery power is low.
• Replace batteries soon
bAtt LowW
• Battery power is too low for
proper operation.
• Replace batteries
5.2 Service Information
If the troubleshooting section does not resolve or describe your problem, contact your
authorized dealer.
6. TECHNICAL DATA
6.1 Specifications
TABLE 6-1.SPECIFICATIONS
Capacity (g)
100 300 150 300 300 500 900 1500 3000 5000 6000
Readability (g)
0.001 0.001 0.01
0.01 0.1 0.1 0.1 0.1 0.1 1 0.1
Linearity (d)
±3 ±5 ±1 ±1 ±1 ±1 ±1 ±1 ±1 ±1 ±2
Repeatability (g)
0.003 0.005 0.01
0.01 0.1 0.1 0.1 0.1 0.1 1 0.1
Span Mass (g)
100 200 100 200 200 300 500 1000 2000 3000 3000
Linearity Masses (g)
50
100
200
300
100
150
200
300
200
300
300
500
500
900
1000
1500
2000
3000
3000
5000
3000
5000
Weighing Units
Gram, Newton
Tare Range
To Capacity by Subtraction
Stabilization Time (sec)
< 3
Operating Temperature Range
50° - 104°F / 10° - 40°C
Operating Humidity Range
30%-90%
Pan Size (mm)
90 dia.
120 dia.
123 x 124