Carolina Hotplate-Stirrer User Manual
Hotplate/stirrer operating manual

Operation Manual
Revised : January 2012
Hotplate/Stirrer
OPERATING MANUAL
This manual can be used for two models. One model,
the “Hotplate Stirrer”, has two knobs on the control panel
for separate control of heating and stirring functions.
The “Hotplate” has a single knob, to control the heating
function..
1.1 Getting Started
Thank you very much for purchasing our Hotplate / Hotplate
Magnetic Stirrer.
Your Hotplate/Magnetic Stirrer has been C E certified and
designed with durability, reliability, and safety in mind. It is
your responsibility to install this instrument in conformance
with local electrical codes. For safe operation, please pay
attention to the alert symbols through the manual.
This manual contains important operating and safety
information. You must carefully read and understand the
contents of this manual prior to operating this instrument.
Warning
Warning alert - possibility of personal injury
Caution
Caution alert - possibility of damage to the
equipment.
Note
Notes alert - pertinent facts and conditions.
Hot Surface
Hot surface alert - possibility of burning injury
by hot instrument surface
Explosive
Explosive alert - possibility of explosion by
high pressure.
1.2 Product Overview
These products are made with an excellent heat-resistant,
anti-corrosive aluminum casting body with a high quality
powder coating.
The ceramic coated and stainless steel top plate provides
excellent chemical resistance, especially against strong acids
or bases.
Unlike traditional hotplate/magnetic stirrers, these units have a
high precision speed control system. (from 60 to 1500rpm)
with a continuously smooth revolution pattern.
The technically advanced PCB board eliminates an electrical
overload and regulates energy distribution. The bearing type
motor maintains gentle revolution over the lifetime of the
equipment.
1.3 Parts and Functions
1.3.1. Control Panel
1 Temperature Controller
ON/OFF and Temperature Control
2 Heater Lamp
Red light Illuminates when heater is on
3 Speed Controller
ON/OFF and revolution speed control
4 Stir Lamp
Green light Illuminates when motor is on
1.3.2. Main Body
1 Power Cord
2 Fuse Holder
3 Standing Rod Position
4 Screw Position
1.4. Operation
Before Operation
1) Place Hotplate/Stirrer on flat and level surface.
2) Plug unit into power outlet. The voltage must correspond
to the voltage listed on the product name-plate
CAUTION : "Hot Surface. Avoid Contact."
The hotplate will remain hot without visual
indication for some time once the power has
been turned off.
Operation
1) Temperature Control
- Turn temperature control knob clockwise, heat light
illuminates.
- Turn temperature control knob to adjust to the desired heat
position.
2) Stirring Speed Control
- Turn Speed control knob clockwise, stir light illuminates.
- Turn speed control knob to adjust to the stirring speed to
the desired position. (Increase speed slowly.)
1.5. Warning
1. Use a properly grounded electrical outlet
with the proper voltage and current
handling capacity as specified on the name-
plate.
2. Disconnect from power supply before
servicing or cleaning.
3. Always place the Hotplate Stirrer on a flat
and level surface.
4. Do not touch the top plate of the instrument
during operation. The hotplate will remain
hot without visual indication for some time
after you have turned the power off.
5. Do not use in the presence of flammable or
combustible materials; fire or explosion
may result. This device contains
components which may ignite such
materials.
6.
_
Replace the top plate immediately if
damaged by scratching or chipping. A
damaged top plate can shatter during use.
7. Do not use metal foil, metal containers,
sand baths or other insulating material on
the hot plate – This can cause damage to
the top plate and shock hazard can exist.
8. Do not remove or modify the grounded
power plug. Use only properly grounded
outlets to avoid shock hazard. The
instrument is not rated for use in
hazardous atmospheres.
9. Use caution when heating volatile samples;
top surface and element can reach the
“Flash Point Temperature”
of many
chemicals. These hot plates are not
explosion proof. Fire or explosion may
result. Unit contains components which
may ignite such materials.
10. Use appropriate hand and eye protection
_
when handling hazardous chemicals.
11.
Refer servicing of this instrument to
_
qualified personnel.
1.6. Trouble Shooting
Check
Trouble
Shooting
Not heating
Or
Not stirring
Check
power
Supply Cord
Plug firmly into
socket
Check Fuse
Replace Fuse
Contact customer service department.
SI Analytics GmbH
Phone: 800.334.5551
Fax: 800.222.7112
Email: [email protected]
/ 66 5001
1.3. Product Specifications
70-1013
70-1012
Type
Hotplate
Hotplate & Stirrer
Speed Range
N/A
60 to 1500 rpm
Temperature Range
80 degrees C to 380 degrees C
Controller
Electronic Solid State Controller
Material
Ceramic Coated Stainless Steel Top Plate
Dimensions
Plate
190 x 190 mm / 7.5 x 7.5 in
Overall
205(W) x 260(D) x 110(H) mm / 8.5(W) x 11(D) x 4.5(H) in
Power Consumption
Max. 500 watt, 3.0 amp
Electric Supply
120 or 220V, 50/60 Hz
1
3
2
4
1
3
2
4