9
user selected. OF display is shown when the count exceeds 999
999 999.
HOLD:
In the FREQUENCY, PERIOD, TOTAL, RPM modes,
measurement in progress is stopped; and the last complete
measurement is displayed. When HOLD is released, a new
measurement begins.
GATE TIME:
Depending on input frequency
< 10 mS----------- Somewhere between 0.9 and 9mS
< 0.1 S------------- Somewhere between 9 and 90mS
< 1 S----------------Somewhere between 90 and 900mS
< 10 S---------------Somewhere between 0.9 and 9S
NOTE:
LAST MEASUREMENT DISPLAY WILL REMAIN FOR 10 SECONDS AFTER
SIGNAL OFF.
DIMENSIONS AND WEIGHT
Dimensions (WxHxD): 9.4 x 3.5 x 10.6” (240 x 90 x 270mm)
Weigh (Approx)t: 5.5lbs. (2.5 kg.)
1-3. Equipment Ratings
AC Input and Fuse Ratings:
AC Input
Fuse
Power Max.
103
126V AC (50/60Hz)
F 0.5A / 250 V
9 W
206
252V AC (50/60Hz)
F 0.2A / 250 V
Plug and Socket:
3 wire ac power plug and 3 wire outlet
Operating Environment:
TEMPERATURE: 0 °C to + 40 °C (Accuracy Specified at 23 °C ± 5 °C)
HUMIDITY:
up to 85% RH (Relative Humidity) to 40