6.4 Instrument Repair Service

Because of the specialized skills and test equipment required for instrument repair

and calibration, many customers prefer to rely upon B&K Precision for this service.

We maintain a network of B&K Precision authorized service agencies for this

purpose. To use this service, even if the instrument is no longer under warranty,

follow the instructions given in the

Warranty Service Instructions

section of this

manual. There is a nominal charge for instruments out of warranty.

7 Error Messages

The following types of errors may occur:

• Self-test errors

• Calibration errors

• Command errors

The first two types of errors are displayed on the voltage display as: Er xx, where xx is

a number. The command errors are sent through the RS232 interface.

7.1 Self-Test Errors

Er 00: EEPROM does not respond
Ee 01
: DAC does not respond
Er 02: CV not high
Er 03: CV not low
Er 04: CC not high
Er 05: CC not low
Er 06: ADC does not respond
Er 07: ADC not ready
Er 08: No reference voltage
Er 09: ADC conversion error
Er 10: ADC AVdd< 3.0V
Er 11: ADC AVdd> 3.6V
Er 12: Calibration constants checksum failed

7.2 Calibration Errors

Zero voltage calibration errors