Preset Mode

Trouble shooting hints

2. Check if the OFF indicator is lit. If so, press the key to turn the output on.
3. Check if the OCP or OVP indicator is lit. If so, set the OVP or OCP value
appropriately.

If the output is disabled
1. Check if the voltage and current values are zero. If set to zero, set the voltage and
current value again.

If keys are disabled

Check the Lock LED. If it is lit, disable the Lock function.

On/Off

On/Off

Press

followed by the

(Preset)

key. The Preset LED will turn on to indicate

that the Preset mode is now active. The most recently selected group number will

automatically be activated. To activate the preset values from a different group, follow

the instructions in “Selecting a group number”. Press one of the

keys to recall one of the corresponding stored Volt/Amp sets assigned in the previous

step. In this mode, the cursor functionality of the A,B,C, D keys is disabled. To exit this

mode, press

then

(Preset)

.

B

B

Shift

Shift

B

B

Shift

Shift

A

A

C

C

B

B

D V/A

D V/A

Storing Voltage/Current sets

Store up to four sets of voltage/current values to the group number assigned in the previous

paragraph. Press

followed by the

(Save)

key. All 4 cursor LEDs

will blink simultaneously. Press one of the

keys to assign one of the 4 available memory location within this group. Proceed

accordingly for the other 3 sets.

A

A

C

C

B

B

D V/A

D V/A

A

A

C D

B

A

Shift