beautypg.com

B&K Precision 1762 - Manual User Manual

Page 42

background image

42

MAINTENANCE




























ADJUSTMENTS

This unit was accurately adjusted at the factory before

shipment. Readjustment is recommended only if repairs have
been made in a circuit affecting adjustment accuracy, or if you
have a reason to believe the unit is out of adjustment. However,
adjustments should be attempted only if a 4-1/2 digit
multimeter with an accuracy of ±0.1% dcv or better is available
(B+K Precision Model 391A or equivalent).


If readjustment is required, use the following procedure. All

references to left and right are correct when facing the front of
the supply. The functions of the adjustments are shown in Table
2 and their locations are shown in Fig. 18.

I. “A” SUPPLY AND “A” METERING ADJUSTMENTS

1. Connect an accurate (±0.1%) external 4-l/2 digit

multimeter to measure the DC voltage at the output
terminals of the “A” SUPPLY.

2. Disengage the INDEP/TRACK mode switch (out) so that

the power supply is in the INDEPendent operating mode.

3. Set the “A” VOLTAGE controls (both Coarse and Fine)

to maximum (fully clockwise).

4. Adjust R6 (“A” SUPPLY +5V REF) on the main circuit

board (located on the right rear side of the supply) for a
reading as close to 30.40 volts (on the multimeter) as
possible.

5. Set the 0-30 V/4-6.5V switch to the 0-30V position and the

Right V/A switch to the V position.

Table 2. Functions of Calibration Adjustments

ADJ

FUNCTION OF ADJ

LOCATION OF ADJ

R6

“A” SUPPLY +5 V REF. MAIN BOARD

R10

“B” SUPPLY +5 V REF. MAIN BOARD

R119

“B” SUPPLY SERIES
TRACKING

MAIN BOARD

R122

4-6.5V A METER & A
LIMIT

MAIN BOARD

R133

4-6.5V V METER

MAIN BOARD

R134

4-6.5V 3.9V REF.

MAIN BOARD

R159

4-6.5V 6.8V REF.

MAIN BOARD

R163

“A” SUPPLY A METER MAIN BOARD

R164

“B” SUPPLY A METER MAIN BOARD

R304L “B” SUPPLY V METER “B” PANEL METER
R304R “A” SUPPLY V METER “A” PANEL METER

1. Adjust R304 (“A” SUPPLY V METER ADJ) on

the “A” panel meter board (located on the right side
of the supply behind the “A”/4-6.5V LED Display)
for a reading of 30.4 volts on the “A”/4-6.5V LED
Display
.

2. Set the “A” Coarse VOLTAGE control for a

reading of approximately 05.0 volts on the “A”/4-
6.5V LED Display
.

This manual is related to the following products: