TrueCPR Device Instructions for Use
Product Specifications
Product Specifications
Mark of conformity to Canadian and US standards
Mark of conformity to ACA standards
Alert sign associated with a Class 2 device under the R&TTE Directive
(Radio and Telecommunications Terminal Equipment)
Manufacturer’s identification number (part number)
Catalog number
Serial number
Date of manufacture
Manufacturer
Authorized EC representative
SPECIFICATION
DESCRIPTION
Power Requirements
Batteries: 2 Duracell nonrechargeable DL123 cells
Operating Time
180 minutes at room temperature with new batteries
Operating Temperature Device can be used at 0° to 40°C (32° to 104°F).
Applied part temperature may reach 43°C (109°F) during
extended use at ambient temperature of 40°C (104°F).
Shelf Life
Provides 30 minutes of CPR feedback after 24 months without
use at 25°C (77°F), starting with new batteries installed.
Storage Temperature
-30° to 70°C (-22° to 158°F)
Relative Humidity
5 to 90%, non-condensing
Altitude
-382 to 4,572 m (-1,253 to 15,000 ft)
Object and Water
Resistance
IP55 per IEC 60529 and EN 1789
Weight
Less than 0.75 kg (1.65 lb) with batteries installed
Ventilation Prompts
2 ventilation prompts every 30 compressions in No Airway mode.
No ventilation prompts in Airway mode.
Compression Depth
Target depth range of 5 to 6 cm (2 to 2.5 in)
Metronome Rate
104.4
±
1 compressions per minute, consistent with AHA and
ERC guidelines.
Model
80596-000003
SYMBOL
DESCRIPTION