beautypg.com

Product specifications – Physio-Control TrueCPR User Manual

Page 22

background image

18

TrueCPR Device Instructions for Use

Product Specifications

Product Specifications

Mark of conformity to Canadian and US standards

Mark of conformity to ACA standards

Alert sign associated with a Class 2 device under the R&TTE Directive
(Radio and Telecommunications Terminal Equipment)

Manufacturer’s identification number (part number)

Catalog number

Serial number

Date of manufacture

Manufacturer

Authorized EC representative

SPECIFICATION

DESCRIPTION

Power Requirements

Batteries: 2 Duracell nonrechargeable DL123 cells

Operating Time

180 minutes at room temperature with new batteries

Operating Temperature Device can be used at 0° to 40°C (32° to 104°F).

Applied part temperature may reach 43°C (109°F) during
extended use at ambient temperature of 40°C (104°F).

Shelf Life

Provides 30 minutes of CPR feedback after 24 months without
use at 25°C (77°F), starting with new batteries installed.

Storage Temperature

-30° to 70°C (-22° to 158°F)

Relative Humidity

5 to 90%, non-condensing

Altitude

-382 to 4,572 m (-1,253 to 15,000 ft)

Object and Water
Resistance

IP55 per IEC 60529 and EN 1789

Weight

Less than 0.75 kg (1.65 lb) with batteries installed

Ventilation Prompts

2 ventilation prompts every 30 compressions in No Airway mode.
No ventilation prompts in Airway mode.

Compression Depth

Target depth range of 5 to 6 cm (2 to 2.5 in)

Metronome Rate

104.4

±

1 compressions per minute, consistent with AHA and

ERC guidelines.

Model

80596-000003

SYMBOL

DESCRIPTION

or

or

YYYY