3.7.4 M
OVING
PATIENT
ON
TILTED
STRETCHER
In situations where conditions do not allow the
chest to remain horizontal:
1.
Turn the
ON/OFF
knob to
Lock
to
pause
compressions (Figure
).
2. As soon as the patient is in horizontal position,
check the position of the suction cup and
adjust if necessary (see Section 3.4).
3. Turn the
ON/OFF
knob to
Active
.
3.7.5 A
MBULANCE
TRANSPORT
Make sure that the patient is fixed when using
LUCAS Chest Compression System during
ambulance transport. The ambulance cot should
have four-point restraints for the patient’s upper
body in addition to the two lower-body restraints.
3.8 C
HANGING
AIR
SOURCES
WARNING – INTERRUPTED COMPRESSIONS
There may be excessive changeover time if pressure
regulators have to be mounted.
Always use reserve cylinders with pre-fitted
pressure regulators.
To change air source:
1.
Set the
ON/OFF
knob to
Adjust
.
2. Close the valve on the used air cylinder.
3. Disconnect the air hose from the air source.
4. Open the valve on the new air source.
5. Attach the air hose to the new air source.
6. Adjust the height of the suction cup on the
patient’s chest, as described previously on
page 17.
7. Turn the
ON/OFF
knob to
Lock
.
8. Turn the
ON/OFF
knob to
Active
to
continue compressions.
1(