beautypg.com

3 assembly, Ssembly – Physio-Control LUCAS User Manual

Page 16

background image

16

100057-00 F, ©J

OLIFE

AB 2007

LUCAS

TM

C

HEST

C

OMPRESSION

S

YSTEM

, I

NSTRUCTIONS

FOR

U

SE

t

y

u

i

3.3 A

SSEMBLY

WARNING – LETHAL PINCH HAZARD

If the air is connected during the assembly

procedure, and the

ON/OFF

-knob is in the

Active

position, the compressions may seriously injure the

patient or create conditions leading to death of the

patient.

Always follow the described sequence of air

connection and assembly.

1.

Take the back plate out of the bag and approach
the patient (Figure

t

).

2. Instruct those carrying out CPR to interrupt

chest compression.

3. Work in a pair, one person on each side of the

patient.

4. Take hold of the patient’s arms. One of the

LUCAS team should support the patient’s head.

5. Lift up the patient’s upper body and lay the

back plate below the armpits. Ensure that the
patient’s arms are outside the back plate

(Figure

y

).

6. Continue manual chest compressions.
7. Take the upper part of LUCAS out of the bag.

Lift it out by holding on to the handles of the

support legs. Ensure that the support legs have
reached their outer position (Figure

u

).

8. Pull up once on the release rings to check that

the claw locks are open.

9. Interrupt manual chest compression.
10. Place the upper part of LUCAS over the

patient’s chest so that the claw locks of the

support legs will engage with the back plate

(Figure

i

).

11. Start by pressing on the support leg nearest to

you and then the one on the other side so the

support legs lock against the back plate.

12. Check by pulling upwards, that both support

legs have locked against the back plate.