3 error symptoms of ph measurement, Error symptoms of conductivity measurement, Error symptoms of ph measurement – Xylem MPP 350 User Manual
Page 23

MPP 350
What to do if...
23
ba75434e02
07/2005
6.2
Error symptoms of conductivity measurement
6.3
Error symptoms of pH measurement
Error symptom
Cause
Remedy
No temperature or con-
ductivity display
– No connection between meter and
sensor
– Cable defective
– Check connection between meter
and sensor
– Return the sensor
Measurement delivers
implausible conductivity
values
– Incorrect cell constant adjusted at
the meter
– Measuring range exceeded
– Contamination in the area of the
electrodes
– Electrodes damaged
– Check / correct the cell constant
– Observe the application range
– Clean the sensor (see
– Return the sensor
Incorrect temperature
display
– The temperature sensor was not
immersed in the measuring
solution
– Temperature sensor defective
– Observe the minimum immersion
depth
– Return the sensor
Error symptom
Cause
Remedy
No pH display
– No connection between meter and
sensor
– Cable defective
– pH electrode defective
– Check connection between meter
and sensor
– Return the sensor
– Replace pH electrode