Maintenance, cleaning, replacement

MPP 350

16

ba75434e02

07/2005

5.3

D.O. module: Changing the electrolyte solution and mem-
brane cap

Caution

Before starting to work with the sensor, please note the G

ENERAL

MAIN

-

TENANCE

INSTRUCTIONS

on page 14.

General

information

WTW delivers the D.O. module ready to use (see section 3). The electrolyte
solution and membrane cap must only be replaced if:

! a calibration error occurs and the membrane is heavily contaminated
! the membrane is damaged
! the electrolyte solution is exhausted.

Changing the electrolyte solution and membrane cap

Unscrew the membrane cap.

Caution: Electrolyte solution!
For disposal of the membrane
cap and electrolyte solution,
see section 5.7.

Rinse the sensor head with
deionized water.

Carefully rub and dry the
counter electrode with a lint-
free paper towel.

Immerse the sensor head in-
cluding the counter electrode
in RL/G cleaning solution.

Allow to react for 1 to 3 min-
utes.

Thoroughly rinse the sensor
head with deionized water.

Water the counter electrode in
deionized water for at least
10 minutes.

RL-G

H O

2