Xylem Oxi 3210 Dissolved oxygen (D.O.) meter User Manual
Page 29

Oxi 3210
Operation
29
ba75795e02
03/2009
4.4.3
Settings for D.O. sensors
(menu or measurement and calibration settings)
Overview
The following settings are possible for D.O. sensors:
z Salinity correction
z Salinity (salinity equivalent)
z Calibration interval
z Automatic Stability control
Settings
The settings are available in the menu for measurement and calibration
settings. To open the settings, display the required parameter in the
measured value display and press the
After completing the settings, switch to the measured value display with
Menu item
Possible
setting
Description
Calibration /
Calibration record
-
Displays the calibration
record of the last calibration.
Calibration /
Calibration interval
1 ... 999 d
Calibration interval for the
D.O. sensor (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measuring window.
Sal correction
On
Off
Manual salt content
correction for concentration
measurements.
Salinity
0.0 ... 70.0
Salinity or salinity equivalent
for the salt content
correction.
This function is only
available for concentration
measurements if the
manual salt content
correction is switched on.
Stability control
On /Off
Switches on or off the
automatic stability control
during measurement (see
section 4.3.3)