Operation

Oxi 3205

28

ba75797e01

03/2009

4.5

Reset

You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.

4.5.1

Resetting the measurement settings

Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.

The following settings for D.O. measurements are reset to the default
settings with the Reset function:

The sensor settings are reset under theReset menu item in the menu
for calibration and measurement settings. To open the settings, display
the required parameter in the measured value display and press the
key.

Setting

Default settings

Cal. interval

14 d

Comparison meas.

Off

Measured parameter

D.O. concentration

Relative slope (S

Rel

)

1.00

Salinity (function)

Off

Temperature unit

°C