The following symbols indicate the result is out of test range:
Result is higher than range
> >
Result is lower than range
< <
6 The ‘Options’ menu at the bottom then offers the choice to:
‘Choose a Test’ - return to the menu of test programs and select another test
‘Read’
- read further sample tubes of the currently selected test
‘Blank’
- re-blank the instrument
‘Follow-On’
- carry out a continuation test if available.
Continuation Tests (Certain Tests Only)
1 Select ‘Follow-On’ and press [OK] during the result display period of the
foregoing test stage. The ‘Insert Sample’ screen will appear.
Place
SAMPLE TUBE in the test chamber, then press [OK].
2 The instrument will take the reading and calculate the result from the
combination of readings (where appropriate). The result will be displayed:
Chlorine-Total / 5
Phot 008
1.50
mg/l Cl
2
Choose a Test Read Blank
Return
3 During the display period, the same options are available as at the end of a
normal test program. Select ‘Return’ from the ‘options menu' to take the
program back to the start of the first stage of a multiple test procedure to
enable further samples to be tested for the same parameters.
Note that some continuation test procedures involve a standing period. The
photometer may switch off automatically during this time. To avoid the
instrument switching off, use the timer function to time any standing period.
The timer will over-ride the auto switch off function.
Favorite Tests List
The four most recently used tests are listed at the top of the 'Choose a Test'
screen for convenience.
8