Zero point calibration – Xylem Oxix Calibration User Manual

Page 3

background image


Item

Oxix

Name:

FW 846003-001
Release date: 12/07 2011
S/N at release date: 97060

Quick Guide for Oxix
Calibration

Version:

04-2012 KJ

EN 5.4 Oxix Quick guide for calibration 1204.doc

3/7

Zero Point Calibration

A zero point check or calibration is conducted as follows:

The sensor is set for a zero point solution. The current measurement is shown in line 2

(0.10mg/l in this example).

Set the desired measuring value on the first line using the up and down arrow keys.

Save the current measurement setup by pressing OK.


The value is saved in the converter and will correct the measurement value of the sensor.
See section Guide to Oxygen-free Solution for Checking Oxix D.O. Sensor Zero Point.