Quick Guide for Oxix
Calibration

Version:

04-2012 KJ

EN 5.4 Oxix Quick guide for calibration 1204.doc

2/7

New Calibration Menu

Valid from firmware version 846003-001, a new calibration menu and method have been
introduced in the Oxix converter.

A factory calibration of both zero point and span has been made at delivery; these calibrations can
be done independently of each other. It is not usually necessary to calibrate. We recommend
checking the zero position every 6 to 12 months.

The calibration is saved in the converter and you will need to conduct a new calibration if the
sensor is moved to another converter.

We recommend conducting a factory reset of the sensor before conducting a new calibration, since
you avoid that any errors in an existing calibration affect the new calibration.

We have also added a menu with the possibility of adjusting salinity levels in the measurement.