Manual – Xylem Expert 700 User Manual
Page 3

Manual
GB Manual Expert 700 0712
As our products are continuously improved, we reserve the right to make
any change in the stated specifications and dimensions without prior notice.
MJK Automation A/S
Byageren 7
DK-2850 Naerum
Denmark
Tel.: +45 45 56 06 56
Fax: +45 45 56 06 46
[email protected]
www.mjk.com
Safety instructions
1: Read this manual carefully.
2: Be aware of the environment on the installation site. Wear necessary protective equipment and follow all current
safety regulations.
3: Do not operate the equipment outside the specified electrical, thermal and mechanical parameters (see
datasheet). Install the device only in media for which the wetted materials have sufficient durability. (See datasheet for
housing material)
Max. supply voltage is 30 VDC.
Hazardous areas
Expert™ 700 Level l Transmitter is NOT approved for explosion hazardous areas.
Repair
1: Repair of THE equipment must only be made by MJK or by a service representative approved by MJK.
Mechanical mounting
Model 700
1: Mount a suitable hook over the desired measuring location. Note the weight of the cable.
2: Remove the inner conical sleeve from the cable fitting and pull the cable through the outer part. Open the inner
sleeve and fit it around the cable at the desired fixation point and press the inner sleeve into place in the outer part.
Secure the cable fitting by pulling the cable downwards.
3: Lower the pressure transmitter into the wellpipe.
Take care not to hit the bottom hard since it may damage the transmitter!
4: If the transmitter is to be used in a wetwell or other locations with turbulence or other disturbance, it is advisable to
install a pipe (min. nominal diameter = 25 mm) to protect the transmitter from bumping into the wall or other
components.
Cable bracket
Protection
pipe
It is very important that minimum 25 mm
of
the pressure transmitter is not being
covered by the pipe!