H-335

SDI-12 Command and Response Protocol 2-13

Example of a H-335 Extended "Zero Offset" command:

Command

Response

Time Values

Description

"aXZO!"

"a0031"

3 sec 1

Zero the UserAngle

Subsequent Command

Response

"aD0"

a+12.000

The new UserAngleOffset

2.13 Extended “XZO” Command

This command is used to automatically zero the gate angle offset. The “aXZO” command

causes the H-335 to make a fresh angle measurement and use this value to automatically update the

UserAngleOffset value to produce a UserAngle reading = 0.00

Make certain the gate is fully closed and resting on the sill. Wait at least 15 minutes or so after

applying +12V power for the sensor temperature to stabilize before issuing the “aXZO” command.

This command should be used only during installation or maintenance.